In Vivo Postoperative Motion of Fixed and Mobile Medial Pivot Knees Under Weight-Bearing Conditions after Cruciate-Sacrificing Total Knee Arthroplasty

Toru Yoshioka1, Nobukazu Okimoto1,2, Mikiya Sawa1, Kei Asano1, Kenji Obayashi1, Makoto Kawasaki3 and Tokifumi Majima4

1Department of Orthopaedic Surgery, Shimura Hospital, Hiroshima, Japan
2Okimoto Clinic, Hiroshima, Japan
3Department of Orthopaedic Surgery, University of Occupational and Environmental Health, Fukuoka, Japan
4Department of Orthopaedic Surgery, Nippon Medical School, Tokyo, Japan


Background: In vivo three-dimensional (3D) motion under weight-bearing conditions was analyzed postoperatively in medial pivot cruciate-substituting (CS) knee systems with fixed and mobile inserts.
Methods: Tibiofemoral knee kinematics during squatting were captured with X-ray fluoroscopy for 4 patients in each cohort. The 3D motion of implants was analyzed with KneeMotion motion analysis software (LEXI Corporation; Tokyo, Japan). In addition, anterior-posterior (AP) movement of the distal-most points and the angle of axial rotation of the femoral component on the tibial component were assessed in both cohorts.
Results: Mean AP movement of the femoral component on the tibial component was 3.8±0.5 mm on the medial side and 9.5±0.5 mm on the lateral side in the cohort with fixed prostheses and 5.9±2.1 mm on the medial side and 10.0±2.5 mm on the lateral side in the cohort with mobile prostheses. The mean angle of axial rotation of the femoral component on the tibial component was 14.4±1.1 degrees and 8.2±2.7 degrees of external rotation for fixed knees and mobile knees, respectively.
Conclusions: Postoperative motion analysis confirmed that fixed and mobile CS implants, which have a similar design, guided medial pivot motion under weight-bearing conditions. However, motion differed between these implant types after mid-flexion: bicondylar rollback after medial pivot motion was noted in the mobile cohort.

J Nippon Med Sch 2023; 90: 103-110

Keywords
total knee arthroplasty, cruciate-substituting implant, medial pivot motion, motion analysis

Correspondence to
Toru Yoshioka, Department of Orthopaedic Surgery, Shimura Hospital, 3-13 Funairimachi, Naka-ku, Hiroshima, Hiroshima 731-0841, Japan
t-yossy@rc4.so-net.ne.jp

Received, April 28, 2022
Accepted, October 19, 2022